IP Indian Journal of Orthodontics and Dentofacial Research (IJODR) open access, peer-reviewed quarterly journal publishing since 2015 and is published under the Khyati Education and Research Foundation (KERF), is registered as a non-profit society (under the society registration act, 1860), Government of India with the vision of various accredited vocational courses in healthcare, education, paramedical, yoga, publication, teaching and research activity, with the aim of faster and better dissemination of knowledge, we will be publishing the more...

  • Article highlights
  • Article tables
  • Article images

Article statistics

Viewed: 888

PDF Downloaded: 733


Get Permission Phalak, Kerudi, Aley, Saxena, Borole, and Sharma: An efficient three dimensional universal guide for accurate placement of mini implants


Introduction

Anchorage refers to the resistance against displacement by anatomical structures and the control of anchorage is one of the main factors for determining the success of orthodontic treatment.1 Conventional means of anchorage system were extra-oral and intra-oral anchorage. Extra oral anchorage was required to negate the forces generated by the intra-oral fixed appliances. Intra-oral anchorage methods were used for 24 hours, but encountered some percentage of anchor loss. Evolution of skeletal anchorage provided “Absolute Anchorage” using dental implants, 2 miniplates 3 and mini implant 4 for fixed appliances which demanded stationary type of anchorage. Titanium mini implant are currently in trend because of their use in various complex tooth movement, minimal anatomical limitations for placement, minimum tissue destruction, low cost and simpler placement. 5 Placement of the mini implant should be in the keratinized gingiva as placement in non-keratinized gingiva leads to formation of hypertrophic tissue formation and inflammation, leading to failure of the mini implant. 6, 7

Success of orthodontic mini implant depends on root proximity of the screw, cortical bone thickness and placement angle. For prosthetic implants, 3mm between the adjacent root and implant surface is recommended for integration and proper health of the tooth. 8 Till date no data is published regarding minimum distance between the mini implant and the adjacent tooth root surface. Hence, mini implant should be maximum away from the adjacent tooth root surface. 5

Precise selection of mini implant is important for its stability. Placement of mini implant is inter-radicular therefore it should be thinner in diameter, but in order to increase the surface area for stability the length of the mini implant is increased. As reported by Poggio et al , the “Safe Zones” for the placement of mini implants in maxilla are mesio-distally between the second and first premolar palatally, and buccopalatally between the first and second premolar. In mandible, mesiodistally the greatest amount of bone was found between first and second premolar; and least at first premolar and canine region. Buccolingually the thickness is more at first and second molars while least at first premolar and canine, thus these sites are suitable for implant placement.9 The thickness of gingiva must also be considered for selection of mini implant, to determine the length of the neck. 10

Increasing the vertical angulation of mini implants by 30o during placement increases the cortical plate contact by 1.5 times than by placing perpendicular to the cortical bone. 11 Increasing this placement angle will increase the contact with the cortical plate. 12 According to Zhao et al, a placement angle of 50o to 70o gave best stability. 13 However, El-Beialy et al statistically reported no significant result with vertical placement angle and the implant success. 14 The stability of the mini implant also increases with age above 18 years and in males as reported by Fayed et al.,15 due to increase in cortical bone thickness.

A 1 mm of clearance of alveolar bone is required for periodontal health of the adjacent tooth from the surface of the mini implant. Thus a minimum of 3 mm of inter-radicular bone is required for the placement of 1 mm diameter of mini implant. 9, 16

Complications of mini implants

  1. Mobility of mini implants

  2. Proximity of tooth root

  3. Oro-antral communication

  4. Peri-implantitis

  5. Undesirable tooth movement

  6. Mini implant fracture

  7. Mini implant slippage 17

  8. Nerve involvement 17

Among the above mentioned complications the proximity of the tooth root to the mini implant surface is utmost important for the stability of mini implant. Hence, the implant must be inserted in a predetermined angulated position and path to avoid root proximity and failure. Several grids and guides have been designed for placement of mini implants, 18, 19, 20 but this appliance is much more accurate for site selection and pre determined angulation of mini implant insertion.

Method of construction

Three dimensional universal guide consists of

A metallic grid for site selection (Figure 1)

The grid was fabricated using 0.7mm round stainless steel wire. The dimension of grid were 10 x 10 mm into which 4 horizontal and 4 vertical wires were welded. A vertical arm through the middle was used for the attachment between the base arch wire and grid. This base arch wire (0.019 x 0.025) passes through the slots of the brackets and tubes.

Implant positioning guide (Figure 2).

Guide is fabricated using a jackscrew, one guiding pin of the jackscrew is cut. A 1 mm wire is used for the formation of the 2 helix guide (internal diameter 6 mm) and transverse arm (length 10 mm), the 2 helices must be unicentric and symmetrical (for confirmation hold the appliance at one hand distance such that when seen from one helix the other helix should not be visible). Solder this appliance to the jackscrew on one side and on other side weld a 0.019 x 0.025 inch base arch wire so as to position the guide. The proximal view shows the central driving shaft and the angulated helices with soldered arch wire (Figure 2b).

Method of placement

The steps for implant placement are as follows,

  1. Initial levelling and alignment must be accomplished for the placement of grid and guide.

  2. Infiltration with local anaesthesia

  3. Place the grid in position with the horizontal base wire in bracket slots securely (Figure 3) and shoot an IOPA for that region with angulation less than 100 from ideal. This angulation was chosen because with ideal angulation the grid was observed more occlusally on IOPA, hence to get higher accuracy for the implant placement site the angulation was modified.

  4. Radiographically, select the proper slot for implant placement (Figure 4), correlating clinically with the mucogingival junction and puncture the selected site with probe(Figure 5). Later remove the grid.

  5. Place the implant guide in position, adjust the guide vertically (by opening the screw or by closing) and horizontally such that the point of puncture is in center of the first helix (Figure 6 ).

  6. Now hold the mini implant in the implant driver and place the implant driver in the guide such that the shank of the implant driver is not touching the second helix. Start inserting the mini implant and keep a watch on the second helix of the guide so that the shaft is not touching the ring or is symmetrical from all the sides (Figure 7).

  7. Insert the implant till the neck and then release the implant driver and remove the guide from position. Tighten the screw, if required, in case the neck is visible (Figure 7).

  8. Take confirmatory radiograph to evaluate the proximity of the implant with the adjacent tooth roots (Figure 8).

The necessary torque incorporation in base arch wires to adapt the grid and guide should be judged clinically, if required.

Advantages

  1. No need for customised fabrication

  2. Can be used in all four quadrants including anterior segments

  3. Grid and guide can be autoclaved.

  4. It is very reliable and accurate method of placement of implants.

Figure 1

Metallic Grid with horizontal base arm and vertical connecting arm

https://s3-us-west-2.amazonaws.com/typeset-prod-media-server/fd485d99-6bc2-470e-bd15-ab09de5a1635image1.png
Figure 2
https://s3-us-west-2.amazonaws.com/typeset-prod-media-server/fd485d99-6bc2-470e-bd15-ab09de5a1635image2.png

Implant placement guide (A) the transverse arm given 150 degree angulation for implant angulation

  1. First helix

  2. Transverse arm

  3. Second helix

  4. Driving shaft

Figure 3

Grid in place for implant site selection

https://s3-us-west-2.amazonaws.com/typeset-prod-media-server/fd485d99-6bc2-470e-bd15-ab09de5a1635image3.png
Figure 4

Radiograph for site selection interproximally

https://s3-us-west-2.amazonaws.com/typeset-prod-media-server/fd485d99-6bc2-470e-bd15-ab09de5a1635image4.png
Figure 5

Punch the selected site with probe

https://s3-us-west-2.amazonaws.com/typeset-prod-media-server/fd485d99-6bc2-470e-bd15-ab09de5a1635image5.png
Figure 6

Guide placed in position and adjusted vertically by closing the screw(A) the punched site in center of the helix (B)

https://s3-us-west-2.amazonaws.com/typeset-prod-media-server/fd485d99-6bc2-470e-bd15-ab09de5a1635image6.png
Figure 7

Implant placed in the center of the helix

https://s3-us-west-2.amazonaws.com/typeset-prod-media-server/fd485d99-6bc2-470e-bd15-ab09de5a1635image7.png
Figure 8

Implant placed with 150 angulation with proper interproximation between the roots.

https://s3-us-west-2.amazonaws.com/typeset-prod-media-server/fd485d99-6bc2-470e-bd15-ab09de5a1635image8.png

Source of Funding

None.

Conflict of Interest

None.

References

1 

W Park J S Park Y M Kim H S Yu K D Kim Orthodontic extrusion of the lower third molar with an orthodontic mini implantOral Surg, Oral Med, Oral Pathol, Oral Radiol Endodontol2010110416

2 

W E Roberts F R Helm K J Marshall R K Gongloff Rigid endosseous implants for orthodontic and orthopedic anchorageAngle Orthod19895924756

3 

Keith H. Sherwood James G. Burch William J. Thompson Closing anterior open bites by intruding molars with titanium miniplate anchorageAm J Orthod Dentofac Orthop20021226593600

4 

H M Kyung Overview development of orthodontic micro-implants for intraoral anchorageJ Clin Orthod2003373218

5 

Shingo Kuroda Kazuyo Yamada Toru Deguchi Takashi Hashimoto Hee-Moon Kyung Teruko Takano Yamamoto Root proximity is a major factor for screw failure in orthodontic anchorageAm J Orthod Dentofac Orthop20071314S68S73

6 

Birte Melsen Carlalberta Verna Miniscrew implants: The Aarhus anchorage systemSemin Orthod20051112431

7 

B G Maino J O Bednar P A Pagin P A Mura The spider screw for skeletal anchorageJ Clin Orthod2003372907

8 

Melanie R. Wood Stanley G. Vermilyea A review of selected dental literature on evidence-based treatment planning for dental implants: Report of the Committee on Research in Fixed Prosthodontics of the Academy of Fixed ProsthodonticsJ Prosthetic Dent20049244762

9 

P M Poggio C Incorvati S Velo A Carano Safe zones”: a guide for miniscrew positioning in the maxillary and mandibular archAngle Orthod2006761917

10 

Renu Parmar Vijay Reddy Satheesh Kumar Reddy Deepti Reddy Determination of soft tissue thickness at orthodontic miniscrew placement sites using ultrasonography for customizing screw selectionAm J Orthod Dentofac Orthop201615046518

11 

C H Moon D G Lee H S Lee J S Im S H Baek Factors associated with the success rate of orthodontic miniscrews placed in the upper and lower posterior buccal regionAngle Orthod2008781016

12 

Shouichi Miyawaki Isao Koyama Masahide Inoue Katsuaki Mishima Toshio Sugahara Teruko Takano-Yamamoto Factors associated with the stability of titanium screws placed in the posterior region for orthodontic anchorageAm J Orthod Dentofac Orthop200312443738

13 

L Zhao Z Xu X Wei Z Zhao Z Yang L Zhang Effect of placement angle on the stability of loaded titanium microscrews: a microcomputed tomographic and biomechanical analysisAm J Orthod Dentofac Orthop2011139562835

14 

Amr Ragab El-Beialy Amr Mohamed Abou-El-Ezz Khaled Hazem Attia Ahmad Mohamed El-Bialy Yehya Ahmed Mostafa Loss of anchorage of miniscrews: A 3-dimensional assessmentAm J Orthod Dentofac Orthop200913657007

15 

Mona Mohamed Salah Fayed Pawel Pazera Christos Katsaros Optimal sites for orthodontic mini-implant placement assessed by cone beam computed tomographyAngle Orthod20108093951

16 

Eric J.W Liou Betty C.J Pai James C.Y Lin Do miniscrews remain stationary under orthodontic forces?Am J Orthod Dentofac Orthop20041261427

17 

Neal D. Kravitz Budi Kusnoto Risks and complications of orthodontic miniscrewsAm J Orthod Dentofac Orthop20071314S43S51

18 

Narendra S Sharma Pushpa V Hazarey RH Kamble Preethi N Sharma Universal Wire Grid for Implant Placement in Three DimensionsWorld J Dent201341746

19 

DilipN Kumar K Mithun Naveen Shamnur KrishnaB. R Gopal A simple and efficient grid for accurate orthodontic implant placementDent Med Res2015326

20 

Raj Kumar Maurya Ankur Gupta Jaishree Garg Chandresh Shukla A custom made grid to guide placement of temporary anchorage devicesAPOS Trends Orthod20155299303



jats-html.xsl


This is an Open Access (OA) journal, and articles are distributed under the terms of the Creative Commons Attribution-NonCommercial-ShareAlike 4.0 License, which allows others to remix, tweak, and build upon the work non-commercially, as long as appropriate credit is given and the new creations are licensed under the identical terms.

Article type

Short Communication


Article page

191-194


Authors Details

Yogesh Subhash Phalak, Veerendra V Kerudi, Mrunal S Aley, Shobhit K Saxena, Prasad A Borole, Madhulika Sharma


Article Metrics


View Article As

 


Downlaod Files